19"" - корпус и переносная версия - ULTRAMAT 23 |
Технические данные
General
|
|
Measured components
|
Max. 4, of which up to 3 infrared-sensitive gases plus oxygen
|
Measuring ranges
|
2 per component
|
Characteristics
|
Linearized
|
Display
|
LCD with LED backlighting and contrast control, function keys 80 characters (4 lines/20 characters)
|
Position of use
|
Front panel vertical
|
Design, enclosure
|
|
Weight
|
Approx. 10 kg
|
Degree of protection, 19" unit and desktop unit
|
IP20 to EN 60529
|
Electrical characteristics
|
|
EMC interference immunity (ElectroMagnetic Compatibility), (safety extra-low voltage (SELV) with safe isolation)
|
According to standard requirements of NAMUR NE21 (08/98) or EN 50081-1, EN 50082-2
|
Power supply
|
100 V AC, +10%/-15%, 50 Hz, 120 V AC, +10%/-15%, 50 Hz, 200 V AC, +10%/-15%, 50 Hz, 230 V AC, +10%/-15%, 50 Hz, 100 V AC, +10%/-15%, 60 Hz, 120 V AC, +10%/-15%, 60 Hz, 230 V AC, +10%/-15%, 60 Hz
|
Power consumption
|
Approx. 60 VA
|
Electric inputs and outputs
|
|
Analog output
|
per component, 0/2/4 to 20 mA, NAMUR, floating, max. load 750 Ω
|
Relay outputs
|
8, with changeover contacts, freely selectable, e.g. for range identification, loading capacity, 24 V AC/DC /1 A, floating, non sparking
|
Binary inputs
|
3, designed for 24 V, floating
|
|
|
|
|
|
|
Serial interface
|
RS 485
|
AUTOCAL function
|
Automatic analyzer calibration with ambient air (depending on measured component), cycle time adjustable from 0 (1) to 24 hours
|
Options
|
Supplementary electronics with 8 additional binary inputs and relay outputs, e.g. for external automatic calibration and for PROFIBUS PA or PROFIBUS DP
|
Ambient conditions
|
|
Permissible ambient temperature
|
|
|
+5 … +45 °C
|
|
-20 … +60 °C
|
Permissible ambient humidity
|
< 90% RH (relative humidity) for storage and transport
|
Permissible pressure variations
|
700 … 1200 hPa
|
Sample gas pressure
|
|
Sample gas pressure
|
|
|
unpressurized
|
|
unpressurized suction mode, factory-adjusted with 2 m hose at sample gas outlet; an upper range value calibration is required in case of other venting conditions
|
Sample gas flow
|
72 … 120 l/h (1.1 … 2 l/min)
|
Sample gas temperature
|
0 … 50 °C
|
Sample gas humidity
|
< 90% RH (relative humidity) no condensation
|
Technical data, infrared channel
|
|
Measuring ranges
|
See Ordering Data
|
Chopper compartment purge
|
Inlet pressure ca 3000 hPa, purging gas consumption ca 100 ml/min
|
Time response
|
|
Warm-up period
|
Approx. 30 min (at ambient temperature), (maximum accuracy is achieved after approx. 2 hours)
|
Response time (T90 time)
|
Dependent on the length of the sample gas cell, the sample line and the programmable damping
|
Damping (electronic time constant)
|
Selectable from 0 to 99.9 s
|
Measuring response
|
|
Noise of output signal
|
< 1 % of smallest measuring range (see rating plate)
|
Display resolution
|
dependent on selected measuring range; the number of digits after the decimal point can be selected
|
Minimum detectable quantity
|
1% of current measuring range
|
Resolution of output signal
|
< 0.1% of output signal span, max. 4.5 digits
|
Characteristic
|
Linearized
|
Linearity error
|
In largest measuring range: < 1% of full-scale value; in smallest measuring range: < 2% of full-scale value
|
Reproducibility
|
≤ 1% of smallest measuring range
|
Influencing variables
|
|
Drift
|
|
|
Negligible
|
|
< 2% of smallest measuring range/week
|
Temperature
|
Max. 2% of smallest possible measuring range according to rating plate per 10 K with an AUTOCAL cycle time of 6 h
|
Atmospheric pressure
|
< 0.2% of measuring range per 1% change in pressure, corrected by internal pressure sensor
|
Power supply
|
< 0.1% of output signal span with a variation of ± 10%
|
Power frequency
|
± 2% of full-scale value with a frequency variation of ± 5%
|
Technical data, oxygen channel
|
|
Measuring ranges
|
0 … 5% or 0 … 25% O2, parameter can be set
|
Typical combustion exhaust gas
|
Influence: < 0.05% O2
|
Lifetime
|
Approx. 2 years with 21% O2; continuous operation < 0.5% O2 will damage the measuring cell
|
Time response
|
|
Response time (T90 time)
|
Dependent on dead time and selectable damping, not > 30 s with a sample gas flow of approx. 1.2 l/min
|
Measuring response
|
|
Noise of output signal
|
< 0.5% of full-scale value
|
Display resolution
|
< 0.2% of full-scale value
|
Resolution of output signal
|
< 0.2% of output signal span
|
Reproducibility
|
≤ 0.05% O2
|
Influencing variables
|
|
Oxygen content
|
operation at times < 0.5 % O2 distorts the measured value
|
Auxiliary gas
|
the oxygen sensor must not be used when the auxiliary gas contains the following components: H2S, chlorine or fluorine compounds, heavy metals, aerosols, mercaptans, alkaline components (as e.g. NH3 in % range)
|
Humidity
|
H2O dew point ≥ 2 °C; the oxygen sensor must not be used with dry sample gases (no condensation)
|
Drift
|
|
|
Negligible
|
|
1 % O2/year in air, typical
|
Temperature
|
< 0.5% O2 per 20 K, referred to a measured value at 20 °C
|
Atmospheric pressure
|
< 0.2% of measured value per 1% pressure variation
|
|
Чертеж
ULTRAMAT 23, 19“ unit, dimensions in mm
ULTRAMAT 23, desktop unit, dimensions in mm |
Схема подключенияPin assignment
ULTRAMAT 23, pin assignment (standard)
ULTRAMAT 23, pin assignment of the optional PROFIBUS interface board
19“ unit
ULTRAMAT 23, 19“ unit, e.g. one infrared component with oxygen measurement
Desktop unit
ULTRAMAT 23, portable unit, in sheet-steel housing, gas and electrical connections
ULTRAMAT 23, designation of the different labels |
Заказные данные
Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
7MB2335-.....-.... | ULTRAMAT23 ГАЗОВЫЙ АНАЛИЗАТОР ДЛЯ ИЗМЕРЕНИЯ 1 ГАЗА И КИСЛОРОДА (ОПЦИЯ) | 12 | по запросу (D) | 1 |
7MB2337-.....-.... | ГАЗОВЫЙ АНАЛИЗАТОР ULTRAMAT 23 ДЛЯ ИЗМЕРЕНИЯ 2 ГАЗОВ И КИСЛОРОДА | 12 | по запросу (D) | 1 |
7MB2338-.....-.... | ГАЗОВЫЙ АНАЛИЗАТОР ULTRAMAT 23 ДЛЯ ИЗМЕРЕНИЯ 3 ГАЗОВ И КИСЛОРОДА | 12 | по запросу (D) | 1 |
|
Аксессуары
Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
A5E00056834 | AUTOCAL BOARD WITH 8 BINARY- INPUTS/OUTPUTS AND PROFIBUS PA | 0.5 | 6 - 8 недель (C) | 1 |
A5E00057159 | AUTOCAL BOARD WITH MIT 8 BINARY INPUTS AND 8 BINARY OUTPUTS AND PROFIBUS DP | 0.5 | по запросу (D) | 1 |
A5E00059149 | SAFETY FILTER | 0.005 | 6 - 8 недель (C) | 1 |
C79000-G5200-C216 | MANUAL (GERMAN) ULTRAMAT 23 | 0.1 | 4 - 6 недель (B) | 1 |
C79000-G5272-C216 | MANUAL (ITALIAN) ULTRAMAT 23 | 0.1 | 4 - 6 недель (B) | 1 |
C79000-G5276-C216 | MANUAL (ENGLISH) ULTRAMAT 23 | 0.1 | 4 - 6 недель (B) | 1 |
C79000-G5277-C216 | MANUAL, ULTRAMAT 23 (FRENCH) | 0.1 | 4 - 6 недель (B) | 1 |
C79000-G5278-C216 | MANUAL (SPANISH) ULTRAMAT 23 | 0.1 | 4 - 6 недель (B) | 1 |
C79451-A3364-D61 | ELAN GATEWAY SIPROM GA UP-DATE COMMUNICATION BY TCP/IP AND ETHERNET | 0.5 | 4 - 6 недель (B) | 1 |
C79451-Z1589-U1 | INTERFACE CONVERTER RS 485 / RS 232 (RETROFIT SET) | 1 | 6 - 8 недель (C) | 1 |
| |
|