Технические данные
Analytical performance
|
|
Measuring range
|
internally adjustable
|
Minimum detection limits (20 °C, 1000 hPa, 1 m path length)
|
gas-depending: HF: 0.1 vpm, HCl: 0.2 vpm, NH3: 0.3 vpm, H2O: 1 vpm, O2: 200 vpm, CO:200 vpm, CO2: 200 vpm
|
Smallest recommended measuring range (20 °C, 1000 hPa, 1 m path length)
|
gas-depending: HF: 0 … 2 vpm, HCl: 0 … 10 vpm, NH3: 0 … 10 vpm, H2O: 0 … 50 vpm
|
in combination with HF, HCl, NH3
|
0 … 5 Vol%, O2: 0 … 1 Vol%, CO: 0 … 1 Vol%, CO2: 0 … 1 Vol%
|
Largest recommended measuring range (20°C, 1000 hPa, 1 m path length or smaller)
|
gas-depending: HF: 0 … 10 Vol%, HCl: 0 … 25 Vol%, NH3: 0 … 100 Vol%, H2O: 0 … 100 Vol%, %, O2: 0 … 100 Vol%, CO: 0 … 100 Vol%, CO2: 0 … 100 Vol%
|
The max. ranges possible with an analyser might be dependent on the measurement conditions and its individual configuration. In case, a required range is larger than 200 times the smallest range given above, please, contact Siemens to discuss your application.
|
General
|
|
Concentration units
|
vpm, Vol%, mg/Nm3 (EU/US)
|
Display
|
Digital concentration display (5 digits with floating decimal point)
|
Mounting orientation
|
Front plate vertical
|
Laser protection class
|
Class 1, safe to the eye
|
Laser power
|
Varies with application
|
Certificates
|
CE mark, ATEX
|
Linearity
|
better than 1%
|
Accuracy
|
better than 2% of reading above the minimum detection limit
|
Design, enclosure
|
|
Protection class
|
IP20 according to EN 60529
|
Dimensions
|
177x440x380 mm
|
Weight
|
approx 13 kg
|
Electrical characteristics
|
|
Power supply
|
100 … 240 V AC 50-60 Hz, automatically adjusted to the system
|
Power consumption
|
50 W
|
EMC-compatibility
|
According to EN 61326 and stan-dard classification of NAMUR NE21
|
Electrical safety
|
According to EN 61010-1, overvoltage classification II
|
Fuse specifications
|
100 ... 240 V: 2.5T/250
|
Time response
|
|
Warm-up temperature at 20 °C ambient
|
Approx. 15 min
|
Delay of display (T90)
|
< 1 s
|
Electrical time constant
|
0.3 (adjustable), typically 1 … 3 s
|
Dead time
|
< 1 s
|
Response time
|
better than 3 s, application dependent
|
Electrical damping
|
1 … 100 s, selectable
|
Measurement response
|
|
Output signal fluctuation
|
2% of measurement value
|
Precision
|
< 2% … < 5% of measurement value, application dependent
|
Zero point drift
|
negligible
|
Drift of measurement value
|
negligible
|
Deviation of linearity
|
< 1% of measurement value
|
Influencing variables
|
|
Ambient temperature
|
< 1%/10 K of measurement value
|
Ambient pressure
|
< 1%/50 hPa
|
Measurement gas pressure
|
< 2% change of reading upon change in pressure of 50 hPa
|
Power supply
|
< 1% at change of output signal range of ± 10%
|
Tilting
|
< 1% for non-horizontal mounting of the sensors below 15°
|
Electrical inputs and outputs
|
|
Number of measurement channels
|
1 … 3, optional
|
Analog output
|
2, 4 … 20 mA, potential-free, ohmic resistance max. 750 Ω
|
Analog inputs
|
2, designed for 4 … 20 mA
|
Binary outputs
|
6, with changeover contacts, configurable, AC/DC 24 V/1 A, potential-free
|
Binary inputs
|
6, designed for 24 V, potential-free, configurable
|
Communication Interface
|
Ethernet 10BaseT (RJ-45)
|
Ambient conditions
|
|
Temperature range
|
+5 … +45 °C during operation, -40 … +70 °C during transportation and storage
|
Humidity
|
< 85% RH, above dew-point
|
Ambient pressure
|
700 … 1200 hPa
|
|
Чертеж
LDS 6, 19“ central unit, dimensions in mm |
Схема подключенияPin assignment
LDS 6, 19“ central unit, pin assignment
Optical and electrical connections
LDS 6, 19“ central unit, optical and electrical connections |
Заказные данные
Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
7MB6121-.....-.... | LDS 6 DIODE LASER SPECTROMETER, CENTRAL UNIT, | 14 | по запросу (X) | 1 |
|
Аксессуары
Заказной № | Описание | Вес (кг) | Доставка | Мин. зак. (шт) |
A5E00534675 | CALIBRATION AND VERIFICATION KIT VERSION 1 | 1.5 | по запросу (X) | 1 |
A5E00823339003 | CALIBRATION KIT RC3009 TUEV/ MCERT, NH3, 2 CELLS, VERSION 1 SERVICE CALIBRATION SET FOR ANALYSERS CERTIFIED ACC. TO 17 TH BIMSCHV. THIS SET IS SUITABLE TO PERFORM REGULAR LINEARITY CHECKS AS REQUIRED FROM THE TEST AUTHORITY | 4.8 | по запросу (X) | 1 |
A5E00823339004 | CALIBRATION KIT RC3009,TUEV/ MCERT, NH3/H2O, 3 CELLS, VERSION 1, SERVICE CALIBRATION SET FOR ANALYSERS CERTIFIED ACC. TO 17 TH BIMSCHV. THIS SET IS SUITABLE TO PERFORM REGULAR LINEARITY CHECKS AS REQUIRED FROM THE TEST AUTHORITY | 5.8 | по запросу (X) | 1 |
A5E00823339005 | CALIBRATION KIT RC3009 TUEV/ MCERT, H2O, 2 CELLS VERSION 1 SERVICE CALIBRATION SET FOR ANALYSERS CERTIFIED ACC. TO 17 TH BIMSCHV. THIS SET IS SUITABLE TO PERFORM REGULAR LINEARITY CHECKS AS REQUIRED FROM THE TEST AUTHORITY | 4.8 | по запросу (X) | 1 |
A5E00823339013 | TUEV/MCERT CALIBRATION KIT NH3, VERSION 2 | 4.8 | по запросу (X) | 1 |
A5E00823339014 | TUEV/MCERT CALIBRATION KIT NH3/H2O VERSION 2 | 5.8 | по запросу (X) | 1 |
A5E00823339015 | TUEV/MCERT CALIBRATION KIT H2O VERSION 2 | 4.8 | по запросу (X) | 1 |
A5E00854188 | LDS6, 3 CHANNEL EXTERNAL POWER SUPPLY | 3 | по запросу (X) | 1 |
A5E01075594 | CALIBRATION VERIFICATION KIT NH3, VERSION 2 | 3.8 | по запросу (X) | 1 |
| |
|